Microsoft word - k061542 fmp.doc

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE

A. 510(k) Number:
B. Purpose for Submission:
De Novo clearance
C. Measurand:
Plasmodium antigens
D. Type of Test:
Qualitative, in vitro immunochromatographic assay E. Applicant:
Binax, Inc., d/b/a Inverness Medical Professional Diagnostics F. Proprietary and Established Names:
Binax NOW® Malaria
G. Regulatory Information:
a) Regulation section:
Plasmodium species antigen detection assay 21 CFR 866.3402. b) Classification:
Class II
Product Code:
OAX
c) Panel:
83 Microbiology
H. Intended Use:

a) Intended use(s):
The Binax NOW® Malaria Test is an in vitro immunochromatographic assay for
the qualitative detection of Plasmodium antigens circulating in human venous and
capillary EDTA whole blood of individuals with signs and symptoms of malarial
infection. The test targets the histidine-rich protein II (HRPII) antigen specific to
Plasmodium falciparum (P.f.) and a pan-malarial antigen, common to all four
malaria species capable of infecting humans - P. falciparum, P. vivax (P.v.), P.
ovale
(P.o.), and P. malariae (P.m.). It is intended to aid in the rapid diagnosis of
human malaria infections and to aid in the differential diagnosis of Plasmodium
falciparum
(P.f.) infections from other less virulent malarial infections. Negative
results must be confirmed by thin / thick smear microscopy.
b) Indication(s) for use:
The Binax NOW® Malaria Test is for the laboratory diagnosis of malaria in
individuals with signs and symptoms consistent with malaria infection.
c) Special condition for use statement(s):
The device is for prescription use only
d) Special instrument Requirements:
NA
I. Device Description:
In vitro immunochromatographic immunoassay J. Substantial Equivalence Information:
K. Standard/Guidance Document Referenced (if applicable):
L. Test Principle:
The Binax NOW® Malaria Test is an immunochromatographic membrane assay that
uses monoclonal antibodies to detect Plasmodium falciparum antigen and pan-malarial
antigen (an antigen shared by all Plasmodium species causing human malaria) in venous
and capillary whole blood specimens. These antibodies, and a control antibody, are
immobilized on a membrane support as three distinct lines and are combined with a
sample pad, which is impregnated with visualizing particles conjugated to control and
anti-malaria antibodies, to create a test strip. This test strip is mounted in a book-shaped,
hinged test device, along with wash and absorbent pads, intended to aid in the clearing of
the membrane when the device is closed.
To perform the test, whole blood is applied to the sample pad. Malarial antigen present in
the sample reacts to bind the anti-malaria conjugated antibody. Reagent A is added to the
bottom of the test strip and allows the antigen-conjugate complexes to migrate along the
test strip, where they are captured by the immobilized antibodies, forming the Test
Line(s). Immobilized control antibody captures control conjugate, forming the Control
Line. Once the blood sample has migrated the length of the test strip, the device is closed,
allowing Reagent A that has been added to the wash pad to clear the test strip of excess
blood.
Test results are interpreted by the presence or absence of visually detectable pink-to-
purple colored lines. A positive test result, read in 15 minutes, will include the detection
of both a Test Line (or Test Lines) and a Control Line. A negative test result, read in 15
minutes, will produce only a Control Line, indicating that malarial antigens were not
detected in the sample. Failure of the Control Line to appear, whether the Test Line(s) is
present or not, indicates an invalid result.
M. Performance Characteristics (if/when applicable):

Analytical performance:
a) Precision/Reproducibility:
A blind study of the Binax NOW® Malaria Test was conducted at 3 separate sites
using panels of blind coded specimens containing negative, limit of detection, and
low positive P.f. and P.v. samples. Participants tested each sample multiple times
on 3 different days. There was 97% (140/144) agreement with expected test
results, with no significant differences within run (replicates tested by one
operator), between run (3 different days), between sites (3 sites), or between
operators (6 operators). The overall percent detection of each sample type is summarized below. Overall Percent Detection of P.f. and P.v. Samples
Sample Type
% Detection
* One operator called a negative sample a P.f. positive. a. Linearity/assay reportable range:
b. Traceability, Stability, Expected values (controls, calibrators, or method):
c. Detection limit:
P.f. and P.v. Limits of Detection:
In the study described above, Binax NOW® test clinical limit of detection (LOD)
for P.f., defined as the parasitemia level in infected blood that produces positive
Binax NOW® test results approximately 95% of the time, was determined to be
1001-1500 parasites per µl and the clinical LOD for P.v. was determined to be
5001-5500 parasites per µl.
d. Analytical specificity:
To determine the analytical specificity of the Binax NOW® Malaria Test, 28 pathogenic microorganisms (7 bacteria, 5 protists and 16 viruses) that may be present in whole blood were tested. All were negative when tested at the concentrations listed below. PATHOGEN TESTED
CONCENTRATION
Interference from Exogenous Blood Components:
The following substances that may be artificially introduced into whole blood were
evaluated in the Binax NOW® Malaria Test at the concentrations listed and were found
not to affect test performance. Note: The analytical effects of these drugs on the Binax
NOW® test were studied by taking whole blood and spiking it with quantities at high
therapeutic concentrations and then testing these samples.
Substance Type
Substance
Concentration
* Doxcycline is also used as an antibiotic, typically at a lower dose than that tested in this study.
Interference from Endogenous Blood Components:
The Binax NOW® Malaria test was evaluated for possible interference from high levels
of endogenous blood components, based on guidelines described in CLSI EP7. EDTA
whole blood samples were tested that contained hemoglobin, protein, bilirubin
(conjugated and unconjugated) or triglycerides at concentrations above physiological
levels. None of the endogenous blood components affected test performance.
Interference from Unrelated Medical Conditions:
To assess the impact of unrelated medical conditions on the specificity of the Binax
NOW® Malaria Test, 116 specimens from subjects with a variety of medical conditions
unrelated to malaria were tested. Only five (5) of the 116 specimens tested produced a
false positive result on the Binax NOW® Test, four (4) from subjects known to be
positive for rheumatoid factor and one (1) from a subject with a positive human anti-
mouse antibody (HAMA) titer.
Comparison studies:
a. Method comparison with predicate device:
b. Matrix comparison:
a. Clinical sensitivity:
Clinical Sample Performance – Binax NOW® Malaria Test Sensitivity
& Specificity – Endemic Population:

The performance of the Binax NOW® test was compared to Giemsa malaria microscopy
in a multi-center prospective study conducted in 2001 outside the U.S., in regions
considered endemic for malaria. A total of 4,122 whole blood specimens collected from
patients presenting with malaria-like symptoms were evaluated on the Binax NOW®
test. Microscopy was considered positive only when asexual malaria forms were
detected, since asexual forms (not gametocytes) are indicative of active infection.
Forty-four percent (1,796/4,122) of the tested population was microscopy positive for
malaria, including 557 patients with P.f., 1,187 with P.v., 16 with P.m., 2 with P.o., and
34 with mixed P.f./P.v. infections. Fifty-nine percent of patients were male, 41% female,
19% pediatric (<18 years) and 81% adult (≥18 years). Binax NOW® test performance
for detection of the individual malaria species and for mixed P.f./P.v. infections is
summarized below.
No differences in Binax NOW® Malaria Test performance were observed based on
patient age or gender. Binax NOW® test specificity for P.f. trends slightly lower
(89.4%) in the 5% of patients who were on anti-malarial drug therapy, than in patients
not receiving therapy (94.4%), but does not achieve statistical significance.

Detection of P.f. Infection

Binax NOW® test sensitivity and specificity for detection of P.f. vs. microscopy is
presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites
per µl) observed in microscopy.
Binax NOW Malaria Test Sensitivity and Specificity for P.f. vs. Microscopy

SENSITIVITY FOR P.f.
Parasitem
SPECIFICITY FOR P.f.
% Specificity
Detection of P.v. Infection Binax NOW® test sensitivity and specificity for detection of P.v. vs. microscopy is presented below. Sensitivity was evaluated based on the levels of parasitemia (parasites per µl) observed in microscopy. There were 68 samples generating two Binax NOW® test lines that were microscopy positive for P.v. only. When these samples are included in the true positive calculation, Binax NOW® test sensitivity for overall detection of P.v. increases from 68.9% to 74.6% (886/1,187). Binax NOW Malaria Test Sensitivity and Specificity for P.v. vs. Microscopy

SENSITIVITY FOR P.v.
Parasitem
SPECIFICITY FOR P.v.
Detection of P.m. and P.o. Infection Binax NOW® test sensitivity was 43.8% (7/16) for detection of P.m. and 50% (1/2) for detection of P.o. When five P.m. microscopy positive samples that generated two test lines in the Binax NOW® test are included in the true positive calculation, Binax NOW®
test sensitivity for P.m. increases from 43.8% to 75.0% (12/16).
Detection of Mixed P.f./P.v. Infection
Thirty four samples were both P.f. and P.v. positive by microscopy, based on the
detection of asexual forms of both species. The Binax NOW® test detected 32 of these
samples by generating both test lines, for a sensitivity of 94.1% (95% CI of 81-98%).
b. Clinical specificity:
The performance of the Binax NOW® test was compared to Giemsa malaria microscopy
in a prospective study conducted in the eastern US in 2006-2007. One hundred (100)
whole blood specimens collected from febrile patients were evaluated on the Binax
NOW® test and on microscopy. All 100 samples were negative for malaria on
microscopy, and 99 of these samples generated negative Binax NOW® test results,
yielding a specificity of 99% (99/100) in this low incidence population. Binax NOW®
test specificity versus microscopy is presented below.
Binax NOW® Malaria Test Specificity vs. Microscopy
+ / - % Spec
P.v., P.o., P.m.


N. Proposed labeling:

The labeling is sufficient and it satisfies the requirement of 21 CFR Part 809.10. WARNING
This test should only be used by laboratories that have or can acquire blood
samples containing Plasmodium falciparum
for use as a positive control. It is
recommended that the level of the positive control used challenge the assay
cutoff.

Clinical performance has not been adequately established for P. ovale (P.o.) and P. malariae (P.m.). The user must establish performance characteristics of this test with these Plasmodium species. The test is not intended for use in screening asymptomatic populations.

O. Conclusion:

The submitted material in this premarket notification is complete and supports a substantial equivalence decision.

Source: http://www.mdic.org.tw/UserFiles/CategoryFile/2011824142937.pdf

Dirkfinal2.pmd

The Millennium Development Goals The evidence is in: deworming helps meet theMillennium Development GoalsSchistosomes and soil-transmitted helminths have Today, control of these infections can be achieved throughfew rivals in terms of prevalence. They occurregular treatment with inexpensive, single-dose andthroughout the developing world, but are mosthighly effective drugs so safe t

Ohne titel

Project on ‘FIELD APPROACH FOR PREVENTION OF DISABILITIES THROUGH COMMUNITY VOLUNTEERS IN RURAL AREAS ADJOINING BOMBAY [PHASE – III]’  [Sponsored by The German Dentist’s Association [HDZ], Germany] Vidnyan Bhavan, 11 VN Purav Marg, Sion-Chunabhatti, Mumbai - 400 022, Maharashtra Tel: 091-22-25223040; Fax: 011-91-22-25296486It is a sad fact that for some years to come millions of ou